General Information

4Gates is an innovative self-treatment device for migraines, based on the simultaneous stimulation of multiple trigger points in the body. Using advanced technology, the device reduces pain and tension, and improves quality of life — without the need for medication.

The 4Gates system is suitable for self-use at home or in the office, as well as for professional therapists as part of broader treatment methods.

Operating 4Gates

(It is recommended to watch the instructional video explaining the process.)

  1. Disconnect the charging cables.

  2. Place the foot units on your feet according to the attached diagram.

  3. Use the pressure adjustment screw on each unit to set the desired pressure level.

    • The pressure should create a mild pain sensation.

    • The sensation is expected to decrease during the session.

    • If the pressure causes discomfort, slightly loosen the adjustment screw.

  4. Repeat the same procedure for the hands, following the attached diagram.

נקודות לחץ

5. Turn on the system by pressing the button.
The treatment duration is 20 minutes.

6. After the session, remove the units.

7. Recharge the units after each use.
It is recommended to leave them connected to the charger.

While Device Use

Guidelines for Optimal Results:

  • Use the 4Gates system in a quiet and comfortable environment, free from conversation or other distractions.

  • The device requires your attention to achieve the best results, so it’s important to avoid any disturbances.

  • Dim the lights and ensure the room temperature is comfortable.

  • Turn off your phone or switch it to silent mode. It is recommended to keep it away from you.

  • Sit on a comfortable chair or sofa.

  • Rest your feet on a wide cushion.

Proper Conduct for Effective Treatment:

  • Close your eyes.

  • Breathing properly during treatment significantly enhances its effectiveness.
    This breathing should be natural — no counting or effort required.

    • Inhale: Breathe in deeply through the nose or mouth.

    • Hold: Retain the air for a brief moment.

    • Exhale: Slowly release all the air through the mouth.

    • Pause: Wait one second before the next inhale.

    • Repeat: Continue this breathing pattern throughout the entire session.

  • Do not rush to remove the device at the end of the session.

Preventive and Pain-Relief Use:

  • Using the device daily enhances the effectiveness of each treatment.

  • The more effective the treatment, the greater the impact on the 4 Gates — reducing pain and even preventing it altogether.

  • It is recommended to reduce and, if possible, avoid the use of painkillers.

Charging

Use the included USB-C hub/splitter to charge each unit of the 4Gates system.

4Gates Model A – While charging, the blue LED light at the top of the device blinks.

4Gates Model B – While charging, the blue LED light at the front of the device blinks.

It is recommended to use the included base during charging, as shown in the image.

Common Issues

Issue: Pressing the button produces no response
Solution: The unit with the button has a depleted battery. Recharge this unit. It is recommended to charge all units together.

Issue: How can the system be reset if needed?
Solution: Disconnecting and reconnecting the power resets each unit. During the reset, the front light of the device will blink in different colors.

Issue: The splitter cable is damaged.
Solution: A new splitter can be ordered from the company website.

Issue: Units are not charging or charging very slowly.
Solution: Use a standard 10W charger. A compatible charger can be purchased on the company website.

Maintenance and Handling

Store the device in a cool, dry place, making sure it is not exposed to direct sunlight or placed near heat sources.

Handle the device gently to prevent damage from drops or impact with hard surfaces.

Avoid contact between the units and water or other liquids, as this may cause damage.

Do not disassemble or modify the units in any way.

The device’s battery should not be replaced by the user.

Store the device out of reach of children or pets to prevent accidental damage or injury.

If any of the units show signs of damage or malfunction, stop using them immediately and contact customer support for guidance and assistance.

User Safety Instructions

Stop using the device immediately if it causes discomfort, pain, or any other adverse reactions.

Do not use the device under the influence of alcohol or drugs.

Consult a doctor before using the device during pregnancy or breastfeeding.

Avoid using the device on injuries, bruises, swelling, or other skin conditions.

If you have been diagnosed with heart conditions, high blood pressure, epilepsy, or any other medical issue, consult a healthcare professional before use.

The device is intended for users aged 18 and above. Keep it out of reach of children.

Device Safety Instructions

Keep the device away from water or moisture.

The device is intended for personal use and suitable for use in a home or office environment, in an area free of dust, sunlight, wind, etc.

If the device is used by more than one person, the plastic parts that come into contact with the skin may be cleaned with a cloth slightly dampened with water.

Only use a certified charger that complies with IEC/EN/UL 62368-1 standards, with an output of 5V DC and a maximum current of 2A.

Only use the charging cable provided by Ramot L.V.